On Power-up, the model M will do the following: 

  1. Flash all three LEDs for 1⁄2 second
  2. Clear the screen
  3. Display the model and firmware revision
  4. Allow the user to change the power level
  5. Allow the user to change the protocol
  6. Clear the screen
  7. Display current power level and protocol
  8. Begin operation: green LED flashes with each pulse


NOTE: Settings can not be changed while operating, only at power-up.
You must shut the power OFF, then back ON to make changes to settings. 

This is to prevent settings from being accidentally changed during operation.


Detailed Power-Up Steps

1.  Start-Up Screen

The first screen you will see is the Start-Up screen.  This screen shows the model number, firmware revision, and the copyright. 

The screen will display for 5 seconds then advance to the next screen automatically unless you press the < SET > button (white button = SET), which advances to the next screen immediately.


2.  Power Adjustment Screen

The next screen is the Power Adjustment screen. This screen shows the power/intensity level as a power bar and number from 1 to 15. Below that it shows the currently selected protocol name (in this case “B5 – C5”). You adjust thepower by pressing the white button < CHANGE >. With each press of the white button, the power will reduce by 1 until it reaches power = 1, then it will jump back and start from 15 again. As before, press the < SET > button (red button = SET), which advances to the next screen immediately, or just wait 5 seconds.



3. Protocol Selection Screen

The next screen is the Protocol Selection screen. This screen shows the currentlyselected protocol name (in this case “B5 – C5”). You can select a different protocolby pressing the white button < CHANGE >. With each press of the white button, the power will move down the list of protocols until it reaches the end, then it will jump back and start from the beginning again. As before, press the < SET > button (red button = SET) to run immediately, or just wait 5 seconds.



4. Running Screen

This is the running screen. The green LED will flash every time a pulse is sent to the coils. To reset the device or change the settings, turn the power OFF, then ON.